RECALIBRATION OF THE ENVIRO-SCAN UNIT ON THE FORMA FREEZER

 

  1. Simultaneously press Battery% and ; should result in a reading of   1

 

  1. Press in the access code: 1, 2, 3, 4

 

  1. Press:    Enter; should result in a reading of   CAL

 

  1. Press:    Control Set Point; should provide a reading of the true internal temperature (this temperature reading should be verified by a reading of a thermometer placed inside the freezer)

 

  1. If the internal freezer temperature does not agree with the Enviro-Scan reading then recalibrate

 

  1. Simultaneously press Battery% and ; should result in a reading of   1

 

  1. Press in the access code: 1, 2, 3, 4

 

  1. Press:    Enter; should result in a reading of   CAL

 

  1. Press:    Enter; will provide the Enviro-Scan temperature reading

 

  1. Hold down either the or the key until the reading corresponds to the control set point temperature

 

  1. Press: Enter; the Enviro-Scan unit should be recalibrated