Amicon Spiral Cartridge Concentrator

Description - Operation - Cleaning - Storage - Sterilization - Testing - Limits


Description

Diaflo spiral cartridges are offered in four MW cut-offs and three sizes. They are especially useful for concentration or desalting (diafiltration) of proteins. They are not recommended for solutions containing suspended particles. Solutions should be pre-filtered (150 mesh, 100 µm) before flow processing. In the spiral cartridge, a "sandwich" of membrane layers is wrapped spirally around a permeate collection tube. Membrane layers in the sandwich are separated by a flow spacer which maintains the retentate flow channels at 0.8 mm (0.03 inch). It also enhances turbulence in the process stream to reduce fouling and gel formation. A cartridge may contain several such "sandwiches", depending on its size. The edges of the membranes are bonded together, so that the process solution cannot enter the flow channel where the fluid permeating the membrane is collected. This fluid moves spirally inside the membrane sandwich to the central permeate collection tube, and then out through the "header" assembly.

MW cut-offs

Solute MW* Membrane Type
Y3 Y10 Y30 Y100


Raffinose
Bacitracin
PVP-15
Cytochrome C
alpha-Chymotrypsinogen
PVP K-30
Albumin
PVP-K60
PVP-K90
Gamma Globulin, Fraction II
Dextran T-500
Apoferritin
IgM

504
1,400
10,000
12,400
24,500
40,000
67,000
160,000
360,000
160,000-900,000
500,000
480,000
960,000
25
>80
-
>98
>98
-
>98
>98
>98
>98
>98
>98
>98
-
20
55
>95
>98
>80
>98
-
-
>98
-
>98
>98
-
-
-
10
>95
60
>98
>90
-
>98
-
>98
>98
0
-
-
0
10
20
20
65
>85
>95
-
>95
>95
At 29 psi (1.9 bar)

Operation

  1. Clean and rinse reservoir and tubing manually.
  2. Using pure glycerin, lubricate O-rings. Do not use silicone or petroleum jelly. Place the two black rubber gaskets at both ends of the cartridge.
  3. Place cartridge outlet side down on a firm surface (necessary to prevent telescoping of the cartridge). Using a slight twist motion, gently push permeate connector tube on inlet header of the ultrafiltration system into collection tube on inlet side of cartridge. Install the outlet header in the same way.
  4. Secure both headers with stainless steel clamps. Do not overtighten clamps.
  5. Fill process reservoir with 2 liters of 0.1 N NaOH. Close the drain valve, open the backpressure valve, direct permeate back to process reservoir. Switch pump to "forward". Turn pump on at mid-speed. Slowly close the backpressure valve until the typical operation pressure of 25-30 psi (1.7 - 2.0 bar) is reached. Recirculate for 10 minutes.
  6. Drain system and repeat recirculation 2 x with dH2O.
  7. Drain system thoroughly. Fill reservoir with (prefiltered !) sample and start concentrating. Ratio can be monitored by the decrease in retentate volume. Never let the sample run completely through. Never let the cartridge run dry.
  8. Monitor the system closely because pressure increases during the process of concentration.
  9. Clean system after use.

 

 

 

For specific cleaning and storage solutions see below.


Cleaning

System Draining and Water Flush

  1. Drain system promptly after use. Do not dry out the membrane.
  2. With the cartridges still installed, flush the system immediately with clean water.
  3. Drain the system completely.

Chemical Recirculation

  1. Refill the system with 2 liters of an approved cleaning solution. Recirculate for 20 to 60 minutes at room temperature. Recirculation at 40°C to 50°C is possible, but cleaning solution with sodium hypochloride may only be used below 23°C.
  2. Leave the permeate line open during operation. Direct the permeate back to the supply reservoir.
  3. Drain the cleaning solution from the system. With permeate ports open, rinse the cartridges with two times with distilled or deionized water for 5 minutes at 25-30 psi (1.7 - 2.0 bar).
  4. Flush system with fresh distilled or deionized water until permeate reaches neutral pH and all traces of cleaner are gone.
  5. To check the effectiveness of the cleaning cycle, measure the distilled or deionized water flux.
  6. To continue operation, fill system with next sample. If no more runs are planned, flush system with storage solution.

Cleaning Solutions

Reagent Final Conc. Volume Mass
dH2O   1800 ml  
NaOH (MW 40.00) 0.1N   8 g

and/or one ore more of the following:

urea (MW 60.06) 6 M   721 g
NaCl (MW 58.44) 1 - 2 M   175 g
lab detergent (e.g. Sparkleen®) 0.1 - 0.3 % 2 - 6 ml  
sodium hypochloride (e.g. Chlorox Bleach: 5.25% by Wt.) 50 - 100 ppm 2 - 4 ml  
nitric acid (HNO3 70%, 15.8 N, for metal ion removal) 0.05 N 6.33 ml  

Adjust volume to 2 liter with dH2O.


Storage


Sterilization


Testing


Limits (environmental resistance)

Return to: Home - LabLinks - GH-PRL-PL